SKC Pocket Pump
The Most Versatile Programmable Low Flow Air Sampling Pump.
The Pocket Pump 20 to 225 ml/min miniature PC-programmable personal sample pump provides up to 12-hour run times for sampling gases/vapours. The pump is ideal for use with sorbent tubes. The Pocket Pump is UL Listed for intrinsic safety.
For more information on the SKC Pocket Pump, contact your local Air-Met office.
- Accommodates typical flow rates for sorbent tubes
- Model available with twin ports for bag sampling
- Only 4.5 x 2.2 x 1.4 inches (11.4 x 5.6 x 3.6cm) and 5 ounces (142 grams)
- Easily fits a shirt pocket or clips on a belt
- Sample with up to 4 tubes simultaneously; select from two pump modes:
- Constant flow mode, requires multiple-tube low flow holder and CPC
- Constant pressure mode, requires multiple-tube low flow holder only
- Program delayed start, timed run, or multiple sampling schedules
- Download sampling data to a PC for record keeping and data analysis
- Simple 3-button keypad does it all
- All sampling parameters are continuously updated
- Low battery and restricted flow shut off
- Pump stores run time data in memory
- Auto-restart attempted after flow fault
- Flow unaffected by outside interferences
- Accurate air volumes and data validity ensured
Large easy-to-read LCD
- Displays sampling parameters, battery status, and flow fault at the touch of a button
- Built-in system automatically senses maximum charge and switches to trickle charge to prolong battery life and prevent damage
- Useful for environmental monitoring or 24-hour sampling without harm to the battery pack
- View battery status before, during, and after sampling
- Rugged case withstands rough handling in the industrial workplace
- UL and cUL Listed
- Model available that meets ATEX Directive 94/9/EC (NiCad battery model only)
- Safe in hazardous locations
- Versatile for all industries
Flow Range in Constant Flow Mode
|
20 to 225 ml/min
|
Accuracy Variance Between LCD Reading and Actual Flow Rate (after calibration) | 20 to 225 ml/min ± 5% |
Constant Flow Compensation Range (inlet only) | 20 to 225 ml/min up to 20 inches water back pressure |
Flow Control Accuracy | ± 5% of set-point in constant flow after calibration |
Pressure Range in Constant Pressure (Multiple-tube Sampling) Mode (inlet only) |
1 to 10 inches of water (1.87 to 18.7 mm Hg) at a maximum flow rate of 200 ml/min for 8 hrs (12 hrs if using NiMH battery) |
Constant Pressure (Multiple-tube Sampling) Mode Accuracy | Pressure reading ± 0.5 in of water (0.25 mm Hg) |
Operating Temperature | 32° to 113°F (0° to 45°C) |
Storage Temperature | -4° to 113°F (-20° to 45°C) |
Charging Temperature | 40° to 100°F (5° to 38°C) |
Operating Humidity | 0 to 95% non-condensing |
Tubing | Requires 1/4-inch ID tubing |
Run Time |
With NiMH: 12 hours at 200 ml/min and 10 inches water back pressure, 12 hours at 100 ml/min up to 20 inches water back pressure Run time is 8 hrs at either setting if using NiCad battery
Varies with flow rate and sampling media in line
Using a charger with AC power: Extend run time
|
Timer Display Range | 1 to 9999 minutes |
Timing Accuracy | ± 1% |
Flow Fault | If the pump is unable to compensate for longer than 15 seconds due to excessive back pressure, the pump enters FLOW FAULT mode. The pump goes into HOLD, the fault icon appears on the display, and the accumulated run time display is frozen and retained. After 5 minutes in flow fault, auto-restart is attempted every 5 minutes until flow is corrected. |
Battery Pack |
Rechargeable NiCad battery pack, 2.4 V x 0.6 Ah or
Rechargeable NiMH battery pack, 2.4 V x 1.0 Ah
|
Charge Time (varies with battery capacity and level of discharge) |
< 6 hrs with Pocket Pump charger (battery attached to pump)
Approximately 16 hrs with charger (battery detached from pump)
|
Battery Charge Level Indicator | Icon displayed at full, mid, and low charge |
Weight | 5 oz (142 gm) |
Dimensions | 4.5 x 2.2 x 1.4 in (11.4 x 5.6 x 3.6 cm) |
Case | Impact-resistant stainless steel with ABS/polycarbonate |
Intrinsically Safe
|
UL and cUL Listed |
RFI/EMI Shielding |
RFI/EMI-shielded case
CE marked
Model 210-1002TX is ATEX approved (NiCad battery model only).
|
- Low flow sampling (gases and vapours)