EPAM-5000 Environmental Particulate Air Monitor
Ideal for Ambient, Environmental and Indoor Air Quality Investigations.
SKU
EPAM5000
The portable HAZ-DUST EPAM-5000 particulate monitor is ideal for ambient, environmental, and indoor air quality investigations. The highly sensitive EPAM-5000 uses light scattering to measure particle concentration and provide real-time determinations and data recordings of airborne particle concentration in mg/m3.
Interchangeable size-selective impactors allow PM10, PM2.5, PM1.0, or TSP monitoring. The unique aerodynamic particle sizing and optional 47-mm in-line filter holder loaded with appropriate filter can provide concurrent gravimetric sampling. Download and store sampling data on a PC for analysis and reporting using included DustComm Pro Software.
For more information on the EPAM-5000 Environmental Particulate Air Monitor, contact your local Air-Met office.
Fast, easy setup
Highly sensitive and accurate
- 0.001 to 20.0 mg/m3 for particles 0.1 to 100 µm
- Optional 0.01 to 200 mg/m3 and 0.10 to 2000 mg/m3 ranges available (specify when ordering)
- Accuracy is ± 10% to filter gravimetric SAE fine test dust
Internal air sample pump for 1 to 4 L/min flows*
- User-programmable start and stop time for unattended data collection
- On-screen user-selectable sampling parameters retained in memory for quick, easy deployment when sampling parameters remain unchanged
- Wireless data transmission option transmits data up to 5 miles
- Improved DustComm Pro PC Software provides:
- Greater flexibility in graphical reports — create and label in new "draw and paint" program
- Basic trend analysis
- Live data feed and logging from wireless data transmission
- Wireless network software option available for Windows and Apple
Unique aerodynamic particle-sizing real-time sensor
- Use optional EPA FRM-style filter holder for gravimetric reference
High correlation to EPA PM10 methods and TEOM®
- Ideal for saturation sampling
Up to 24-hour operation on one battery charge
- Continuous monitoring with AC adapter
Immediate display and data storage
Audible alarm siren
Easy-to-use DustComm Pro data analysis and reporting software included
Portable and compact
- 14 x 6 x 10 inches ( 35.6 x 15.2 x 25.4 cm)
- 10 pounds (4.5 kg)
Interchangeable size-selective impactors for EPA parameters
- PM10, PM2.5, PM1.0, or TSP
Pump has a variable flow rate designed for use at 4 L/min to achieve proper size selection.
Sensing Range
|
0.001 to 20 mg/m3
0.01 to 200 mg/m3 (optional); specify when ordering
0.10 to 2000 mg/m3 (optional); specify when ordering
|
Particle Size Range | 0.1 to 100 µm |
Precision | ± 0.003 mg/m3 (3 µg/m3) |
Accuracy | ± 10% to filter gravimetric SAE fine test dust |
Sampling Flow Rate | 1 to 4 L/min with variable adjustment |
Filter Cassette Holder (optional accessory) |
47 mm, FRM Style
37 mm or 25 mm (optional)
|
Alarm Output | 90 dB at 3 ft |
Analogue Output | 0 to 2 vdc |
Recording Time | 1 sec to 15 mos |
Sampling Rate | 1 sec, 10 sec, 1 min, and 30 min |
Data Storage | 21,600 data points |
Security Code | 4-digit combinations |
Memory and Time Storage | > 5 years |
Data Display | Concentration in mg/m3 and TWA, MAX, MIN, STEL, Date, and Time |
Digital Output | RS-232 |
Operating Temperature | 14° to 122°F (-10° to 50°C) |
Storage Temperature | -4° to 158°F (-20° to 70°C) |
Power | Rechargeable lead-acid and interchangeable battery |
Operating Time | > 24 hrs on battery |
Charging Time | Approx. 22 hrs |
Dimensions (case) | 14 x 6 x 10 in ( 35.6 x 15.2 x 25.4 cm) |
Weight | 10 lbs (4.5 kg) |
- Quantifying off-site particulate migration
- Surveying for PM2.5 and PM10
- Monitoring dust generation during drilling and excavation
- Evaluating dust suppression and engineering controls
- Locating and identifying "hot spots"
- Emergency response and fugitive emissions compliance
- Determining level of respiratory protection
- Survey of workplace for OSHA/EPA compliance
- Evaluating worker exposure to airborne contaminants
- Dust collector/ventilation system checks
- Monitoring lung-damaging particles in factories and buildings
- Complementary instrument for all EPA and OSHA personal and ambient particulate air sampling procedures