Defender Series Calibrators
Accurate, reliable and portable primary calibrators for the fast calibration of air sampling pumps.
Turn the Defender on. Connect it to your pump. Take measurements. It’s that simple. Instantly calibrate your gas flow using Quick Start – there is no preparation involved, no messy soap bubble solutions. Defender’s many user-selectable options makes customizing your calibrations easy, giving you exactly what you need in the format you prefer. All you have to do is choose between the 510, the 520 or the 530.
Backed by Proven DryCal® Technology for fast,1% of reading calibrations of personal sampling pumps, the Defender 510, the enhanced Defender 520, and the standardising Defender 530, give you the flexibility to do as much as you need while you work to defend lives.
For use with non-corrosive, non-condensing, non-combustible gases, less than 70% humidity.
For more information on the Defender Series Calibrators, contact your local Air-Met office.
3 Models, 3 Flow Standards
Defender 510 | Volumetric Primary Flow Standard
- Flow Range:
-
Low: 5-500ccm
-
Medium: 50-5,000ccm
-
High: 300-30,000ccm
-
- Single, Continuous and Burst reading options
- Accuracy Volumetric (% of reading) 1%
- Volumetric Flow Readings – mL/min, L/min, cc/min, cf/min
- Data Port: Serial (RS-232)
- 1/4” Barbed Plastic Compression Fittings
Defender 520 | Enhanced Volumetric Primary Flow Standard
- Flow Range:
- Low: 5-500ccm
- Medium: 50-5,000ccm
- High: 300-30,000ccm
-
Single, Continuous and Burst reading options
-
Accuracy Volumetric (% of reading) 1%
-
Temperature & Pressure Sensors in flow stream
-
Temperature Readings: C°, F°
-
Pressure Readings: mmHg, PSI, kPa
-
Volumetric Flow Readings – mL/min, L/min, cc/min, cf/min
-
Data Port: Serial (RS-232)
-
1/4” Barbed Plastic Compression Fittings
- Flow Range:
- Low: 5-500ccm
- Medium: 50-5,000ccm
- High: 300-30,000ccm
-
Single, Continuous and Burst reading options
-
Accuracy Volumetric (% of reading) 1%
-
Accuracy Standardised (% of Reading) 1.2%
-
Temperature & Pressure Sensors in flow stream
-
Temperature Readings: C°, F°
-
Pressure Readings: mmHg, PSI, kPa
-
Volumetric Flow Readings – mL/min, L/min, cc/min, cf/min
-
Standard Flow Readings – smL/min, sL/min, sccm, scf/min
-
Data Port: Serial (RS-232)
-
1/4” Barbed Plastic Compression Fittings
Defender 500 Series | |
---|---|
Approximate Timer Per Reading
|
1-15 seconds
|
Reading Styles (with averaging function user-selectable from 1 to 100 measurements) |
Single (manual), Continuous or Burst |
Gas Compatibility | Non-corrosive gases, non-condesing less than 70% humidity |
Flow Modes | Pressure or Suction |
Measurement Cell Style | Integrated |
AC Adapter/Charger | 12V DC, > 250ma, 2.5mm centre positive |
Battery System | 6V rechargeable, sealed lead-acid, 6-8 hrs typical operation |
Battery Operational Time (5 cycles/min) | 3 hrs backlight on, 8 hrs backlight off |
Pressure and Suction Fittings | 1/4" ID hose barb |
Warranty | 1 yr (battery 6 months) |
Storage Temperature |
0-70°C
|
Ambient Humidity | 0-70%, non-condensing |
Operating Pressure (absolute) | Atmospheric, max 15 PSIA |
Display
|
Backlit graphical LCD |
Data Port (for PC interface) | 1 metre Defender to PC included |
Protective Case | Pelican case available |
Dimensions (H x W x D) | 140 x 140 x 75mm |
Weight | 820g |
Complaince | RoHS CE Compliant |
Defender 510 | Defender 520 | Defender 530 | ||
---|---|---|---|---|
Model and Flow Ranges | 510L, 5-550ccm 510M, 50-5000 ccm 510H, 300-30,000 ccm |
520L, 5-500 ccm 520M, 50-5000 ccm 520H, 300-30,000 ccm |
530L, 5-500ccm 530M, 50-5000 ccm 530H, 300-30,000 ccm |
|
Volumetric Accuracy | 1% | 1% | 1% | |
Standardised Accuracy | n/a | n/a | 1.2% | |
User Selectable Measurement Units | Volumetric | mL/min, L/min, cc/min, cf/min | mL/min, L/min, cc/min, cf/min | mL/min, L/min, cc/min, cf/min |
Standardised | n/a | n/a | smL/min, sL/min, scc/min, scf/min | |
Temperature | n/a | °C, °F | °C, °F | |
Pressure | n/a | mmHg, PSI, kPa | mmHg, PSI, kPa | |
Temperature and Pressure Sensor | No | Yes, in flow stream | Yes, in flow stream | |
Pressure Accuracy | n/a | +/- 3.5mmHG (typical) +/- 7.0 mm (max) |
+/- 3.5mmHG (typical) +/- 7.0 mm (max) |
|
Temperature Accuracy | n/a | +/- 0.8°C (typical) +/- 1.3°C (max) |
+/- 0.8°C (typical) +/- 1.3°C (max) |
- Industrial hygiene
- Environmental applications
- Laboratory applications
- Manufacturing and quality control
- Environmental and ambient air monitoring
- MFC/MFM calibrations
- Semicondictor processing